HomeContactPrint

Turbidity meter

Turbidity meter Type AS82.2

Technical data
Supply voltage:24VAC 50 - 60 Hz / DC / -10%..50%
Power consumption:Approx. 9VA (AC) / 6 VA (DC)
Emitted measurement signal:880nm
Ambient temperature:+5°C .. 45°C
Analogue output:0..20mA or 4..20mA
Relay breaking capacitymax. 250V, 1A, 100VA
Measuring range:>10E 10 cells per ml or >200g per liter

Solid suspension (product dependent)

Calibration required!

PDF Datasheet AS82.2

Turbidity meter Type ATM 100 (incl. Ex-version)

Technical data
Ex-classification:II 2G EEx emb II T4 (pre-amplifier)
Supply voltage:24VAC/DC -20%..+25%
Power consumption:max. 15VA
Number of measuring points:up to max. 4 measuring points at the same time
Ambient temperature:0C … 60C
Analogue output:0…20mA or 4…20mA
Relay breaking capacity:2 per measuring point max. 42 VAC, max. 0.5A, max. 50VA
Measuring range:0 .. 10000NTU <+/- 2% linearized
Display:full graphic display 240x160 pixels monochrom

PDF Datasheet ATM 100

ATM 1

Technical data
Supply voltage:24VAC/DC -20%..+25%
Power consumption:max. 15VA
Number of measuring points:1 measuring point
Ambient temperature:0C … 60C
Analogue output:0…20mA or 4…20mA
Relay breaking capacity:Switch contact C, NC, NO
Measuring range:0 .. 4000 NTU automatically selectable after calibration
Display:Alphanummerical display 2x16 digits